| Model Number |
H7493808015320 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter in two pieces.The balloon was loosely folded.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube was completely separated 58.5cm from the hub.The hypotube fracture surfaces were ovaled and torn, which suggests the device was kinked prior to separation.There are numerous hypotube kinks.Inspection of the remainder of the device presented no other damage or irregularities.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 22-aug-2018.It was reported that crossing difficulties were encountered.The 90% stenosed, 3.25mm x 15mm target lesion was located in the moderately tortuous and calcified right coronary artery.A 3.25mm x 15mm quantum¿ maverick¿ balloon catheter was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed a hypotube break.
|
| |
|