Model Number ACTC1525
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, there was an alarm problem and before the start of the operation for a liver tumor, the radio frequency machine passed the self-test and the rf needle was inserted.When the ablation was ready, the temperature window showed a rapid increase in temperature.Being that high, the device automatically shut down after hh, the operation was successful after replacing the rf needle.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: (street 1, mfr city, region, postal code), evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.The investigation found the device to function normally and within specifications.The water circulated normally through the device.The device read the temperature properly.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL-TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7884907
Report Number1717344-2018-01233
Device Sequence Number16771387
Product Code GEI
UDI-Device Identifier10884524001203
UDI-Public10884524001203
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K973297
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2021
Device Model NumberACTC1525
Device Catalogue NumberACTC1525
Device Lot Number80190139X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Initial Date Received by Manufacturer 08/27/2018
Supplement Date Received by Manufacturer11/30/2018
Initial Report FDA Received Date09/18/2018
Supplement Report FDA Received Date12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-