Model Number UNIVERS REVS CUP,42/NEUTRAL CTD
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 08/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation revealed both the screw and the raised side of the screw hole on the cup had fractured.The fragments were not received for evaluation.The opposite side of the fractured hole was also damaged.In addition, there were scratches and tool marks on both inner and outer surfaces plus a visible wear pattern was observed.The device met all material specifications as received.A root cause for the event cannot be determined at this time.
 
Event or Problem Description
It was reported that the patient required a revision surgery.Original date of surgery was (b)(6) 2015.Per the rep, the ar-9502-42cpc broke in half in the patient when he was opening a bottle of water.The revision was (b)(6) 2018.Parts removed were ar-9502-42cpc (lot 12.714), ar-9550-12 (lot 13.490) and a ¿liner¿ (no part # found per the rep).These were replaced with ar-9502f-42cpc and ar-9550-15.The original cup was done at the 135 degree slot; the revision was done at the 155 degree slot.Original surgery was done in (b)(6), revision in (b)(6) (different surgeons).Additional information obtained 9/13/18: the sales rep has confirmed that, per the surgeon's decision, the distal part of a screw, which is part of the universal reverse cup (ar-9502-42cpc), was stuck in the 135 degree hole of the universal reverse stem that was not explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS REVS CUP,42/NEUTRAL CTD
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key7884924
Report Number1220246-2018-00645
Device Sequence Number228746
Product Code PHX
UDI-Device Identifier00888867061187
UDI-Public00888867061187
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K142863
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberUNIVERS REVS CUP,42/NEUTRAL CTD
Device Catalogue NumberAR-9502-42CPC
Device Lot Number12.714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2018
Initial Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-