This device was received and visually inspected.Visual observation revealed that the tip of the jaw was chipped, preventing a proper closure when the device is actuated.This reported condition could be confirmed.We cannot definitively discern a root cause for this failure mode at this time but typically, broken components are associated with excessive force applied to the device during usage.A review of the device history record indicated that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.Further, a review into the depuy synthes mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
It was reported by the affiliate in (b)(6) that the arthro grasper pen device was broken.According to the reporter, the device would not close properly.It was not reported if there was a delay in the surgical procedure or if a spare device was available for use.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|