• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US ARTHRO GRASPER PEN 45 RIGHT *EA; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ARTHRO GRASPER PEN 45 RIGHT *EA; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Catalog Number 214603
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device was received and visually inspected.Visual observation revealed that the tip of the jaw was chipped, preventing a proper closure when the device is actuated.This reported condition could be confirmed.We cannot definitively discern a root cause for this failure mode at this time but typically, broken components are associated with excessive force applied to the device during usage.A review of the device history record indicated that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.Further, a review into the depuy synthes mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the affiliate in (b)(6) that the arthro grasper pen device was broken.According to the reporter, the device would not close properly.It was not reported if there was a delay in the surgical procedure or if a spare device was available for use.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7885388
Report Number1221934-2018-51567
Device Sequence Number14134607
Product Code NBH
UDI-Device Identifier10886705003860
UDI-Public10886705003860
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214603
Device Lot Number10R01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2016
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/06/2016
Initial Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-