• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER SONICISION; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER SONICISION; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number SCD48
Medical Device Problem Code Intermittent Energy Output (4025)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Evaluation summary: one device was returned for evaluation.The returned product did not meet specification as received.Visual inspection of the disposable hand piece revealed that the waveguide had fractured and the tip had broken off.The broken piece was not returned with the device.The reported condition was confirmed.The waveguide was inspected under magnification to identify the point of initial contact that caused the fracture and eventual break.Investigation personnel concluded that the titanium waveguide was in use when it fractured.Further investigation revealed that the waveguide was excessively torqued to the side during use causing it to come in contact with the moving jaw of the device while it was being activated.This contact caused the waveguide to crack and eventually break off.The investigation identified the root cause of the reported event to be user error.The device instructions for use currently contain a warning against contact between the dissector tip and metal objects (hemostats, clips, staples, retractors, etc.) during activation.This issue is specific to ultrasonic dissectors.The instructions for use (ifu) notes that device users should visually inspect all system components for breaks, cracks, nicks, or other damages prior to use.Ifu also states: do not use damaged components.Use of damaged components may result in injury to the patient or user.No update to the risk management file is required at this time.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the hand piece was being used and stopped working during case, giving red indication.The battery and generator was exchanged with no improvement.The hand piece was replaced and worked properly with original battery/generator.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICISION
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7885401
Report Number1717344-2018-01238
Device Sequence Number2725954
Product Code LFL
UDI-Device Identifier10884521195226
UDI-Public10884521195226
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K141371
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/10/2023
Device Model NumberSCD48
Device Catalogue NumberSCD48
Device Lot Number81240144X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2018
Initial Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age56 YR
-
-