| Catalog Number |
9733449 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.Manufacturing date was unavailable at the time of filing.The suction was returned to the manufacturer.Functional testing and a visual and physical examination found that the component was damaged.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that straight suction was not fully seated in the tray.It was tested and it couldn¿t be verified.It verified once after about a minute and it was about 3mm off.There was no patient present when this issue was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.If information is provided in the future, a supplemental report will be issued.
|
| |
|