• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MOSAIC AORTIC BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MOSAIC AORTIC BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 305C221
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Occlusion (1984)
Date of Event 08/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during the implant of this 21mm aortic bioprosthetic valve, the 21mm aortic bioprosthetic valve was seated and secured into the annulus.The surgeon noted that the left and right coronary ostia were partially blocked and the valve was determined to be oversized.The valve was subsequently explanted and replaced with a 19mm valve of the same model.No additional adverse patient effects were reported. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC AORTIC BIOPROSTHETIC HEART VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key7885490
Report Number2025587-2018-02445
Device Sequence Number2488829
Product Code LWR
UDI-Device Identifier00643169594685
UDI-Public00643169594685
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P990064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/13/2023
Device Model Number305C221
Device Catalogue Number305C221
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2018
Initial Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient Weight78
-
-