• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. SCEPTER XC OCCLUSION BALLOON (4X11 MM); BALLOON CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. SCEPTER XC OCCLUSION BALLOON (4X11 MM); BALLOON CATHETER Back to Search Results
Model Number BC0411XC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The balloon was returned for evaluation.A pinhole was identified in the balloon material at the distal balloon marker band; however, the root cause of the rupture could not be determined.
 
Event or Problem Description
It was reported that the balloon "ruptured" after being inflated once.There was no reported patient injury or intervention.The patient was reported to be doing okay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCEPTER XC OCCLUSION BALLOON (4X11 MM)
Common Device Name
BALLOON CATHETER
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7885998
Report Number2032493-2018-00186
Device Sequence Number1819520
Product Code MJN
UDI-Device Identifier00842636020591
UDI-Public(01)00842636020591(11)180222(17)210131(10)180222123
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113698
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2021
Device Model NumberBC0411XC
Device Lot Number180222123
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2018
Initial Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-