| Model Number |
BC0411XC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The balloon was returned for evaluation.A pinhole was identified in the balloon material at the distal balloon marker band; however, the root cause of the rupture could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon "ruptured" after being inflated once.There was no reported patient injury or intervention.The patient was reported to be doing okay.
|
| |
|