| Model Number |
72202019 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the surgery, the screw fell out when implanting.No patient injury was reported.Back-up device was use to completed the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
One 2.0 minitac ti anchor was returned for evaluation.Visual assessment of the device showed the anchor and suture are free from the inserter.The needles are free from the suture and were not returned.The anchors threads are heavily damaged making dimensional assessment prohibitive.The insertion device shows no damage.With the limited clinical details provided a root cause cannot be determined.Corrected information.
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery, the screw fell out when implanting.No patient injury was reported.There was no backup device avalable, the surgery was complete with a competitor device.
|
| |
|