| Model Number |
BA25-90/I16-30 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in the event will not be returned for evaluation, they remain implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Patient was initially implanted with an afx bifurcated, one vela suprarenal, and one ovation limb.Recently, it was noted that the patient has a possible endoleak type iiib with fabric billowing/tear.Patient is currently being monitored, there has not been any plans for intervention.As of this report, there has not been any additional patient sequelae reported.
|
| |
|
Additional Manufacturer Narrative
|
|
At the completion of the clinical assessment, clinical was able to refute the reported event of a type 3b endoleak, rather there is the fabric billowing in the acceptable range.No device malfunction noted therefore this is no longer a reportable event.
|
| |
|
Event or Problem Description
|
|
At the completion of the clinical assessment, clinical was able to refute the reported event of a type 3b endoleak, rather there is the fabric billowing in the acceptable range.No device malfunction noted therefore this is no longer a reportable event.
|
| |
|