| Catalog Number |
0279351250 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the probe was continuously activating.The procedure was completed successfully with no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: per sales rep, turned it on and it wouldn't turn off.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the detachment of the suction tube to the suction shaft causing a leak internally which creates a short.Inadequate gluing operation on the suction tube and suction adapter, the user accidentally submerges the device in saline.The product was returned for investigation and the failure mode was confirmed and will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the probe was continuously activating.The procedure was completed successfully with no impact to the patient.
|
| |
|