BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM07040
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records were provided to the manufacturer; however, four medical images were provided and are pending review.The lot number for the device was provided.The device history records are currently under review.The device is not available for return.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that after a successful endovascular stent graft placement procedure to cover a tight stenosis, the health care provider ran a post visual image and alleged it appeared there was a vessel dissection at the distal end of the stent graft.Therefore, a second endovascular stent graft was placed to cover the alleged perforation and complete the procedure.Reportedly, hemostasis was achieved, and the patient was fine post procedure.
 
Additional Manufacturer Narrative
Manufacturing review: based on the information available and the investigation performed, there is no indication that the product failed to meet its specification prior to shipment.The lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented, that could have led to the event reported.Investigation summary: a physical sample was not returned.However, images were provided for evaluation.Based on the evaluation of the images provided, it was confirmed that a radiopaque marker was protruding sideways.In addition an accumulation of contrast medium outside the regular course of the vessel is displayed.Based on the images provided a manufacturing related issue could not be confirmed.Based on the information available and the evaluation of the provided images, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the currently valid labeling it was found that the instructions for use (ifu) sufficiently address the potential risk.The ifu states: "the safety and effectiveness of the device when placed across a tight bend including the terminal cephalic arch or across the elbow joint has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." furthermore, the ifu states: "post dilate the stent graft with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.".
 
Event or Problem Description
It was reported that after a successful endovascular stent graft placement procedure to cover a tight stenosis, the health care provider ran a post visual image and alleged it appeared there was a vessel dissection at the distal end of the stent graft.Therefore, a second endovascular stent graft was placed to cover the alleged perforation and complete the procedure.Reportedly, hemostasis was achieved, and the patient was fine post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7887757
Report Number2020394-2018-01741
Device Sequence Number1908452
Product Code PFV
UDI-Device Identifier04049519008431
UDI-Public(01)04049519008431
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFEM07040
Device Catalogue NumberFEM07040
Device Lot NumberANBN2456
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/24/2018
Supplement Date Received by Manufacturer11/07/2018
Initial Report FDA Received Date09/18/2018
Supplement Report FDA Received Date11/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
-
-