• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CLYDESDALE PTC¿ SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CLYDESDALE PTC¿ SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 4986045
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with spinal kyphoscoliosis and underwent olif (oblique lumbar interbody fusion) surgery at l1/2 levels.During surgery, while inserting the cage by approaching from the left side between lumbar l1/2, the cage protruded to the contralateral side and could not be pulled back.So, the doctor had to approach from the right side, and remove the cage from the right side, and changed the direction of the cage and re-inserted it from the right side.There was a delay of more than 60 minute in overall procedure time as a result of this event.There were no patient symptoms or complications as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The physician confirmed that the cage advanced deeper when it was hammered.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The level was l1/2, and the rib and the inserter interfered with each other, and the insertion angle was inclined.So, it was in the state that the space between the vertebral body end plate and the cage became tight and the sleeve was difficult to be removed; thus, it was considered that the sleeve did not come off but the cage was inserted deep when the sleeve remover was used in the above state.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLYDESDALE PTC¿ SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key7888262
Report Number1030489-2018-01253
Device Sequence Number1758492
Product Code MAX
UDI-Device Identifier00643169190733
UDI-Public00643169190733
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133205
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4986045
Device Lot NumberH5405861
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/24/2018
Supplement Date Received by Manufacturer09/20/2018
10/11/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date10/02/2018
10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
Patient Weight69
-
-