• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT EC8+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT EC8+ CARTRIDGE Back to Search Results
Catalog Number 600-9007-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Information (3190)
Date of Event 08/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat ec8+ cartridges that yielded suspected discrepant results on a patient.There was no patient information available at the time of this report.(b)(6).There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.It is suspected that the patient sample was potentially hemolyzed.However, there is not enough information to support the cause.The investigation is underway.
 
Additional Manufacturer Narrative
Apoc incident # (b)(4).The investigation was completed on 10/16/2018.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ac (product complaint level 2 and level 3 investigation procedure).No deficiency has been determined for the lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT EC8+ CARTRIDGE
Common Device Name
EC8+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key7888762
Report Number2245578-2018-00240
Device Sequence Number1908754
Product Code CHL
UDI-Device Identifier10054749000071
UDI-Public10054749000071
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K940918
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2018
Device Catalogue Number600-9007-25
Device Lot NumberH18119
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer10/16/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-