| Model Number |
C501 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer stated that they received an erroneous result for one patient sample tested for ise indirect k for gen.2 on a cobas 6000 c (501) module.The customer also complained of a difference in the serum index values of the sample when repeated.The complained sample is lipemic.The sample initially resulted with a k value of 5.50 mmol/l and this value was reported outside of the laboratory.The sample was repeated, resulting with a k value of 3.99 mmol/l.No adverse events were alleged to have occurred with the patient.The k electrode lot number and expiration date were asked for, but not provided.Calibration and controls are within specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer has had no further erroneous results.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|