Batch review performed on 18 september 2018.Lot 153636: (b)(4) items manufactured and released on 23 september 2015.Expiration date: 2020-06-30 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed on 18 september 2018 by medical affairs director: the clinical evaluation cannot find out what the reason for this problem could have been.No clinical cause seems to be possible.Thread dimensions and hole diameter are standardized.The diameter of the drill could not be verified after the operation, but there is no reason to suspect an error.Metrological analysis performed by quality control manager on 13 september 2018.The dimensional control, has not identified any anomaly or production defect.The device are conforming to the specifications of project.Visual inspection performed on 18 september 2018 by r and d product manager: after dimensional control, no dimensional defects emerge on the screws 01.26.65.35 cancellous bone screw flat head d.6.5 mm l35 mm lot.153636.The cause of the failure of the implant positioning in the related seat is not related to the implant.
|