| Catalog Number |
367981 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that black foreign matter was in the gel of the bd vacutainer® sst¿ blood collection tubes.There was no report of exposure, injury or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received samples and photos from the customer facility for investigation.The photos were evaluated and the customer¿s indicated failure mode for fm with the incident lot was observed.Additionally, evaluation/testing of the customer samples was performed and fm was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Based on evaluation of the customer photos, the customer¿s indicated failure mode for fm with the incident lot was observed.Additionally, evaluation/testing of the customer samples was conducted and fm was observed.Based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.As a result, bd is reviewing specific areas in the manufacturing process where the cause of this issue may have originated.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that black foreign matter was in the gel of the bd vacutainer® sst¿ blood collection tubes.There was no report of exposure, injury, or medical intervention.
|
| |
|