| Model Number |
EUP2015X |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A euphora balloon catheter was attempted to be used to treat a moderately tortuous, moderately calcified lesion with 95% stenosis in the proximal-mid lad.There was no damage noted to the device packaging.There were no issues noted when removing the device from the hoop.The device was inspected with no issues.Negative prep was performed with no issues.The device was not kinked and re-straightened during use.Two balloons were inside the guide catheter, the euphora balloon and an unknown brand balloon.It was reported that the devices became entangled and could not be advanced.When an attempt was made to remove the balloon, a detachment occurred.The euphora balloon separated from the delivery system.All components stayed in the guide catheter and were removed together when the guide was removed.No components remain in the patient.It is also reported that the catheter became stretched.No other devices were tried after they removed the guide.No patient injury is reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The guide catheter was non-medtronic.A second euphora balloon was in the guide catheter and was entangled with the euphora that had the detachment.The second euphora was not damaged.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: deformation was evident to the hypotube from 2cm to 54cm distal to the strain relief.There was a detachment on the transitional tubing 3cm distal to the proximal end of the transition tubing.The support wire was exposed distal to the detachment site.The transitional material was oval and jagged on both sides of the detachment site.The detached portion of the device was loaded on to a guidewire.The distal shaft was entangled around the guidewire.The guidewire lumen appeared deformed and bunched, the inflation lumen appeared wrinkled.A portion of the inflation lumen appeared to be torn.Deformation was evident to the transitional tubing.No deformation was evident to the guidewire.If information is provided in the future, a supplemental report will be issued.
|
| |
|