| Model Number |
EGIA30AV |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the second firing of segmental resection of lung and thoracoscopy procedure, the device could not be fired even an opening and closing check was done without problem.A new device was used to complete the firing without any problems.The event occurred during the procedure but the product was not used for patient.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the reload was pre-fired and engaged in interlock.There were no staples protruding from proximal staple cartridge channel.Functionally the reload was loaded into a post market vigilance instrument, the interlock was overridden, and the reload was applied to test media with proper staple placement and media transection.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time by ceasing the placement of staples and tissue transection and prevent patient harm.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|