| Catalog Number |
MFR181647 |
| Medical Device Problem Code |
Unraveled Material (1664)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Information regarding patient age, gender, weight, medical history, race, and ethnicity was not provided.Procode: krd/hcg.Physical manufacturer name: codman and shurtleff inc., dba depuy synthes products, inc.((b)(4)).Concomitant med products due to character limitation: cook sheath introducer, asahi guidewire, cook guiding catheter, stryker microcatheter, stryker balloon catheter, enpower dcb2 detachment control box.[complaint conclusion]: as reported by a healthcare professional, during a balloon-assisted coil embolization of a splenic artery aneurysm, the 16mm x 47cm micrusframe18 (mfr181647/l10828) thermo-mechanical coil was being placed in the aneurysm as the first coil in the case, but the coil moved due to rapid blood flow in the parent artery and the coil became unraveled.The coil and the stryker microcatheter were removed as a unit and the coil was replaced.The procedure was completed successfully without further incident or delay.No patient complications occurred as a result of the event.The complaint product was new and stored per labeling instructions.The procedure was conducted in accordance with the instructions for use (ifu) and a constant flush was maintained.No visible product defect/damage was noted prior to the event.No unintended detachment was observed in the vessel or in the microcatheter.A cook sheath introducer, asahi guidewire, cook guiding catheter, stryker balloon catheter, and an enpower dcb2 detachment control box were also utilized for the case.No further information was provided.The micrusframe18 was not returned for investigation.A review of manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.With the limited information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the manufacturing process of the unit.Positioning difficulty and coil unraveled/stretched are known potential issues associated with the use of the device.The ifu contains several cautions relating to this situation, including instructions for troubleshooting the situation should it be encountered during use.Assignment of root cause for the event remains speculative and inconclusive, based on the minimal information provided; however, it is possible that aneurysm/vessel characteristics, including the rapid blood flow in the parent artery, and procedural/handling factors, including device interaction and device manipulation, may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Event or Problem Description
|
|
As reported by a healthcare professional, during a balloon-assisted coil embolization of a splenic artery aneurysm, the 16mm x 47cm micrusframe18 (mfr181647/l10828) thermo-mechanical coil was being placed in the aneurysm as the first coil in the case, but the coil moved due to rapid blood flow in the parent artery and the coil became unraveled.The coil and the stryker microcatheter were removed as a unit and the coil was replaced.The procedure was completed successfully without further incident or delay.No patient complications occurred as a result of the event.The complaint product was new and stored per labeling instructions.The procedure was conducted in accordance with the instructions for use (ifu) and a constant flush was maintained.No visible product defect/damage was noted prior to the event.No unintended detachment was observed in the vessel or in the microcatheter.A cook sheath introducer, asahi guidewire, cook guiding catheter, stryker balloon catheter, and an enpower dcb2 detachment control box were also utilized for the case.The product was discarded and is thus not available for evaluation.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Event: additional information received indicated that there was difficulty positioning the coil in the target aneurysm due to rapid blood flow in the parent artery.The coil was removed successfully from the patient without need for additional intervention.The coil delivery system was prepped without any difficulty.No visible product damage was noted prior to use.Excessive force was not applied during the case.The event did not result in a clinically significant delay in the procedure.Initial reporter phone: (b)(6).Complaint conclusion - updated: as reported by a healthcare professional, during a balloon-assisted coil embolization of a splenic artery aneurysm, the 16mm x 47cm micrusframe18 (mfr181647/l10828) thermo-mechanical coil was being placed in the aneurysm as the first coil, but the user encountered difficulty positioning the coil in the aneurysm due to rapid blood flow in the parent artery and the coil moved and became unraveled.The coil and the stryker microcatheter were removed as a unit and the coil was replaced.The coil was removed successfully from the patient without need for additional intervention.The procedure was completed successfully without further incident or delay.No patient complications occurred as a result of the event.The complaint product was new and stored per labeling instructions.The coil delivery system was prepped without any difficulty.The procedure was conducted in accordance with the instructions for use (ifu) and a constant flush was maintained.No visible product defect/damage was noted prior to use.No unintended detachment was observed in the vessel or in the microcatheter.A cook sheath introducer, asahi guidewire, cook guiding catheter, stryker balloon catheter, and an enpower dcb2 detachment control box were also utilized for the case.The user did not apply excessive force at any time during the procedure.No further information was provided.The micrusframe18 was not returned for investigation.A review of manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.With the limited information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the manufacturing process of the unit.Positioning difficulty and coil unraveled/stretched are known potential issues associated with the use of the device.The ifu contains several cautions relating to this situation, including instructions for troubleshooting the situation should it be encountered during use.Assignment of root cause for the event remains speculative and inconclusive, based on the minimal information provided; however, it is possible that aneurysm/vessel characteristics, including the rapid blood flow in the parent artery, and procedural/handling factors, including device interaction and device manipulation, may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Search Alerts/Recalls
|
|