STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HAR9F
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed evidence of clinical use.The tip of the distal end of the blade was broken off and was not returned with the device.Some damage was observed on the surface of the teflon pad where the break occurred.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: - jaws/blade subassembly damage.- incidental and prolonged activation against solid surfaces, such as bone or plastic.The instructions for use (ifu) state: - avoid contact with any and all metal or plastic instruments or objects when the instrument is activated.Contact with staples, clips, or other instruments while the instrument is activated may result in cracked or broken blades, which may be identified by generator solid tone or instrument error.- care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this happens, there may be a system failure signaled by a continuous tone or alert screen when either of the foot pedals or hand control buttons is depressed.- scratches on the blade may lead to premature blade failure.- the entire exposed blade tip and any exposed blade shaft is active and will cut/coagulate tissue when the harmonic focus®+ shears blade is activated.Be careful to avoid inadvertent contact between all exposed blade surfaces and surrounding tissue when using the harmonic focus + instrument.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported for a harmonic scalpel (har9f) that the metal shear broke off during case.The piece was successfully retrieved from the patient without requiring an x-ray nor larger incision.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
MDR Report Key7889734
Report Number0001056128-2018-00058
Device Sequence Number2725967
Product Code NLQ
UDI-Device Identifier07613327348248
UDI-Public07613327348248
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K170456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/20/2021
Device Model NumberHAR9F
Device Catalogue NumberHAR9FRR
Device Lot Number8124491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2018
Initial Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-