• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC APPROACH HYDRO ST MICROWIRE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC APPROACH HYDRO ST MICROWIRE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Catalog Number HMW-14-300-ST
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/18/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field for section was not populated due to a technical limitation.Section product code: dqx this mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.Blank fields on this form that have not been previously submitted indicate the information is unknown or unavailable.
 
Event or Problem Description
The complainant reported that the approach hydro st microwire wire guide was used to cross a calcified occlusion in the patient's foot during a tibial angioplasty procedure.The wire stretched and unraveled upon removal.All of the wire was removed from the patient.It was further noted that the distal weld connection separated.The product problem did not result in any reported serious injuries to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPROACH HYDRO ST MICROWIRE WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7889799
Report Number1820334-2018-02886
Device Sequence Number169932
Product Code DQX
UDI-Device Identifier00827002529397
UDI-Public(01)00827002529397(17)170301(10)4956584
Combination Product (Y/N)N
Initial Reporter CountryAU
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2017
Device Catalogue NumberHMW-14-300-ST
Device Lot Number4956584
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2018
Initial Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
-
-