• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC SAFE-T-J AMPLATZ EXTRA-STIFF SUPPORT HEPARIN COATED WIRE GUIDE; DQX WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC SAFE-T-J AMPLATZ EXTRA-STIFF SUPPORT HEPARIN COATED WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Catalog Number THSCF-35-260-3-AES-SGH-BH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/28/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field for section was not populated due to a technical limitation.Section product code: dqx.This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
The complainant reported that the safe-t-j amplatz extra-stiff support heparin coated wire guide was used during the bav (balloon aortic valvuloplasty) procedure.The physician noted an issue after applying torque and inserting the device tip inside of the left ventricle.The device was removed from the patient, and it was noted that the core wire protruded from the outer spring coil.Another device with the same specifications was used to complete the procedure without any problem.It was further noted that the distal weld connection separated from the coil and safety wire, but the product problem did not result in any reported serious injuries to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE-T-J AMPLATZ EXTRA-STIFF SUPPORT HEPARIN COATED WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7889897
Report Number1820334-2018-02888
Device Sequence Number19262959
Product Code DQX
UDI-Device Identifier00827002116849
UDI-Public(01)00827002116849(17)200106(10)5508640
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/06/2020
Device Catalogue NumberTHSCF-35-260-3-AES-SGH-BH
Device Lot Number5508640
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2018
Initial Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-