STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL 13.7 -11.0
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Code Available (3191)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.7 mm micl 13.7 , -11.0 diopter, implantable collamer lens into the patients right eye (od) and removed it on (b)(6) 2018.During initial surgery due to observation of excessive vault and significant reduction of irido-corneal angles.On (b)(6) 2018 the lens was exchanged for a shorter length lens.The exchange resolved the problem.Additional information reported; patient has residual astigmatism with plans for prk (photorefractive keratectomy) in 1 month.
 
Additional Manufacturer Narrative
Corrected data: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7890067
Report Number2023826-2018-01348
Device Sequence Number16825719
Product Code MTA
UDI-Device Identifier00841542103589
UDI-Public00841542103589
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Model NumberMICL 13.7 -11.0
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer11/01/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date11/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INJECTOR#UNK,LOT#UNK,CARTRIDGE#SPFC-45,LOT#UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
-
-