• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; INTRODUCER, CATHETER Back to Search Results
Model Number 4253540-02
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2016018 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as event two of b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.Additional attempts te receive the sample are being made.A follow-up report will be submitted when the results of the investigation are available.
 
Event or Problem Description
As reported by user facility: event 2.The safety device did not deploy.
 
Additional Manufacturer Narrative
Exemption number e2016018.B.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as event two of b.Braun medical inc.Internal report number (b)(4).The actual device involved in the reported incident was not returned, however a photo was provided for evaluation.This product is subjected to a 100% inspection by an in-line vision system that is calibrated and subjected to regular verifications to ensure it is functioning properly.The defect observed in the provided photo would have been detected and rejected by the in-line vision system.Without the actual sample a thorough sample analysis could not be performed and no specific conclusions can be drawn.Review of the device history record performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.Based on the results of the investigation, no conclusions can be made regarding the cause of the reported event.If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key7890658
Report Number9610825-2018-00197
Device Sequence Number1909137
Product Code DYB
UDI-Device Identifier04046963318246
UDI-Public(01)04046963318246
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K021094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/26/2018
Report Date (Section F) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2023
Device Model Number4253540-02
Device Catalogue Number4253540-02
Device Lot Number18B13G8317
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/17/2018
Device Age7 MO
Event Location Hospital
Date Report to Manufacturer11/26/2018
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer09/17/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date11/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA11/26/2018
-
-