MAUDE Adverse Event Report: STRYKER GMBH SUPERIOR PLATE - DECREASED CURVATURE VARIAX CLAVICLE 6 HOLE / RIGHT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH SUPERIOR PLATE - DECREASED CURVATURE VARIAX CLAVICLE 6 HOLE / RIGHT; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 628026
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
Primary procedure, right clavicle.It was reported that in implanting the variax2 core 3.5 locking screws into the clavicular plate, the screws would not lock - they kept spinning when they should have locked.Rep obtained locking screws from a variax 1 ankle solution set, which locked as expected.The plate is implanted and the reported screws are available for return.Rep reported that surgeon was following the surgical technique throughout the procedure.There was a surgical delay of 5 minutes.
 
Event or Problem Description
Primary procedure, right clavicle.It was reported that in implanting the variax2 core 3.5 locking screws into the clavicular plate, the screws would not lock - they kept spinning when they should have locked.Rep obtained locking screws from a variax 1 ankle solution set, which locked as expected.The plate is implanted and the reported screws are available for return.Rep reported that surgeon was following the surgical technique throughout the procedure.There was a surgical delay of 5 minutes.
 
Additional Manufacturer Narrative
This device is concomitant and did not contribute to the reported failure.Therefore, this complaint is closed without further investigation.If any additional information is provided indicating otherwise the record will be reopened and the investigation will be reworked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERIOR PLATE - DECREASED CURVATURE VARIAX CLAVICLE 6 HOLE / RIGHT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key7891404
Report Number0008031020-2018-00613
Device Sequence Number1682080
Product Code HRS
UDI-Device Identifier07613252441922
UDI-Public07613252441922
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K113760
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number628026
Device Lot NumberY03508
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2018
Initial Date Received by Manufacturer 08/22/2018
Supplement Date Received by Manufacturer10/17/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date11/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age14 YR
-
-