| Catalog Number |
9733450 |
| Medical Device Problem Code |
Imprecision (1307)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device not yet returned to manufacturer.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess).The issue occurred intraoperatively while verifying instruments and delayed the surgery by less than one hour.It was reported that there was difficulty registering the 70-degree suction.The suction did eventually get registered and it was the only instrument showing inaccurate.The surgery was completed with navigation and there was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
The curved suction was returned to the manufacturer for analysis.Analysis found that the returned curved suction was not able to verify when attached to a known good tracker returning a divot error of 3.6 mm to 3.8 mm.Analysis found that the reported event was related to a mechanical issue.If information is provided in the future, a supplemental report will be issued.
|
| |
|