• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) 70 DEG CURVED SUCTION, AXIEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) 70 DEG CURVED SUCTION, AXIEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9733450
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device not yet returned to manufacturer.
 
Event or Problem Description
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess).The issue occurred intraoperatively while verifying instruments and delayed the surgery by less than one hour.It was reported that there was difficulty registering the 70-degree suction.The suction did eventually get registered and it was the only instrument showing inaccurate.The surgery was completed with navigation and there was no impact on patient outcome.
 
Additional Manufacturer Narrative
The curved suction was returned to the manufacturer for analysis.Analysis found that the returned curved suction was not able to verify when attached to a known good tracker returning a divot error of 3.6 mm to 3.8 mm.Analysis found that the reported event was related to a mechanical issue.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
70 DEG CURVED SUCTION, AXIEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key7891845
Report Number1723170-2018-04857
Device Sequence Number19424486
Product Code HAW
UDI-Device Identifier00763000053390
UDI-Public00763000053390
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9733450
Device Lot Number26337
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2018
Supplement Date Received by Manufacturer08/23/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date10/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age9 YR
Patient Weight35
-
-