Model Number 752-7545-11A
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Death (1802); Pain (1994); Patient Problem/Medical Problem (2688)
Date of Event 03/01/2018
Type of Reportable Event Death
Event or Problem Description
On (b)(6) 2018, a revision surgery was performed to extend existing l4-l5 pedicle screw fixation to l3-s1.The surgery also involved vertebroplasty of l5 together with l5-s1 tlif using another manufacturer's cage.During surgery, the patient's blood pressure began to drop and the patient expired.The report indicated the patient's left iliac artery was most likely perforated when pushing the rod onto the screw.The surgeon stated this incident was due to a technical error and not the pedicle screw.Two pedicle screws were in use but it is unknown which screw was involved so two identical medwatch reports are being filed, one for each part number.This is report 1 of 2.
 
Additional Manufacturer Narrative
This follow up report #1 corrects 2 errors in the initial report.Manufacture date corrected from 30-nov-2017 in the initial report to 18-sep-2017 in this follow up.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAGODA
Common Device Name
PEDICLE SCREW
Manufacturer (Section D)
ORTHO DEVELOPMENT
12187 s. business park drive
draper UT 84020
MDR Report Key7891960
Report Number1722511-2018-00032
Device Sequence Number2726222
Product Code NKB
UDI-Device Identifier00822409022519
UDI-Public(01)00822409022519(17)220810(10)A180776
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131785
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/10/2022
Device Model Number752-7545-11A
Device Lot NumberA180776
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/22/2018
Supplement Date Received by Manufacturer08/22/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age75 YR
-
-