• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY Back to Search Results
Catalog Number PNML6F088804M
Medical Device Problem Codes Break (1069); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the distal tip of the neuron max was fractured approximately 82.0 cm from the hub and the packaging mandrel was unsecured from the tape.Conclusions: evaluation of the returned neuron max revealed that the distal tip of fractured.This type of damage typically occurs during removal of the device from the packaging.If the packaging mandrel becomes unsecured from the packaging card it may slip into the packaging tube.This may cause the distal tip of the neuron max to become pinned between the mandrel and packaging tube.If the neuron max is subsequently retracting, the distal atraumatic tip may become damaged.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a medical procedure using a neuron max 6f 088 long sheath (neuron max).The physician removed the neuron max from the sterile package and noticed that a piece of the tip had broken off.The damage to the neuron max was found prior to use.Therefore, the neuron max was not used in the procedure.The procedure was completed using a new neuron max.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX 6F 088 LONG SHEATH
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7891966
Report Number3005168196-2018-01838
Device Sequence Number2365411
Product Code DQY
UDI-Device Identifier00814548012209
UDI-Public00814548012209
Combination Product (Y/N)Y
Initial Reporter CountrySZ
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/21/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2021
Device Catalogue NumberPNML6F088804M
Device Lot NumberF81802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2018
Initial Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-