MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAH
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that they experienced high blood glucose.Customer's blood glucose was over 600 mg/dl.Customer reported that the insulin pump had frozen display as well as insulin pump alarmed after battery changed.Customer also reported that they were unable to clear the alarm.Customer stated that an alarm occurred during that time the buttons stopped.Screen was not blank after inserting new battery.Troubleshoot was performed.Customer advised to insulin pump will need to be replaced and revert to back up plan.Insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump received no button response (act) due to corroded keypad traces.Insulin pump time clock advances no frozen screen noted.The information provided in the section of concomitant product with the initial report was incorrect.The correct information has been included with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7892388
Report Number2032227-2018-30647
Device Sequence Number1434359
Product Code OYC
UDI-Device Identifier00643169446397
UDI-Public(01)00643169446397
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAH
Device Catalogue NumberMMT-523NAH
Device Lot NumberB2523NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2018
Initial Date Received by Manufacturer 09/02/2018
Supplement Date Received by Manufacturer12/04/2018
Initial Report FDA Received Date09/19/2018
Supplement Report FDA Received Date12/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRN RESERVOIRUNOMEDICAL INFST
Outcome Attributed to Adverse Event Other;
Patient Age11 YR
Patient Weight61
-
-