• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1755K PUMP 630G 3ML BK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1755K PUMP 630G 3ML BK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1755K
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 09/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had a critical pump error.The customer¿s blood glucose was 382 mg/dl.The customer was treated with manual injection.Troubleshooting was performed.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device was received with critical pump error (open book image) alarm during testing due to moisture damage on electronic assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT MMT-1755K PUMP 630G 3ML BK MG
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7892624
Report Number2032227-2018-30852
Device Sequence Number19309990
Product Code OZO
UDI-Device Identifier00643169752726
UDI-Public(01)00643169752726
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-1755K
Device Catalogue NumberMMT-1755K
Device Lot NumberHG261A4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2020
Initial Date Received by Manufacturer 09/01/2018
Supplement Date Received by Manufacturer04/24/2020
Initial Report FDA Received Date09/20/2018
Supplement Report FDA Received Date04/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-