| Catalog Number |
413400 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer in the united states notified biomérieux of a false resistant result associated with vitek® 2 ast-gn69 test kit (reference 413400).The customer reported obtaining an incorrect cap survey result where piperacillin/tazobactam was incorrect.The expected result was sensitive.The original result was submitted as resistant >=128 (r).The customer pulled the specimen and performed repeat testing on vitek® 2 and microscan.Vitek® 2 yielded the same result of >=128 (r) and microscan had a result of <=16 (s).The customer mentioned that this portion of the survey is ungraded.There is no patient involvement due to being a cap survey.An internal biomérieux investigation will be initiated.
|
| |
|
Event or Problem Description
|
|
A customer in the united states notified biomérieux of a false resistant result associated with vitek® 2 ast-gn69 test kit (reference 413400).The customer reported obtaining an incorrect cap survey result where piperacillin/tazobactam was incorrect.The expected result was sensitive.The original result was submitted as resistant >=128 (r).The customer pulled the specimen and performed repeat testing on vitek® 2 and microscan.Vitek® 2 yielded the same result of >=128 (r) and microscan had a result of <=16 (s).The customer mentioned that this portion of the survey is ungraded.There is no patient involvement due to being a cap survey.An internal biomérieux investigation will be initiated.
|
| |
|
Additional Manufacturer Narrative
|
|
A customer in the united states notified biomérieux of false resistant results for cap survey strain (d-08) when testing with the vitek® 2 ast-gn69 card.The alcaligenes faecalis ssp.Faecalis strain resulted in resistant mics >/= 128 for piperacillin/tazobactam (tzp).The expected result was susceptible.The strain was submitted for investigation.The customer's submitted strain and the biomerieux cap d-08 survey strains were subcultured on tsab agar, identifications confirmed as a.Faecalis ssp faecalis, and testing included ast-gn69 cards from customer lots (5890758403, 5890708203) and a random lot (5890711103) as well as broth microdilution (bmd) as the reference method for the tzp formulation on this card (tzp03n).911703 (customer's d-08 strain): all three cards tested gave a resistant mic >/= 128.Bmd gave a susceptible mic = 4, demonstrating a major error for the vitek 2.911704 (biomerieux internal d-08 strain): on both customer lots of ast-gn69 cards, susceptible mic = 4 and on the random lot a resistant mic >/= 128 were obtained.Bmd gave a susceptible mic = 4.Card and reference method are in agreement for the customer lots, but the random lot mic results demonstrate a major error for vitek 2 when compared to reference.This is an atypical strain.
|
| |
|
Search Alerts/Recalls
|