Model Number 507
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complaint catheter was returned in a biohazard bag.There is a longitudinal tear in the balloon.The proximal end of the balloon is attached to the outer tubing.The distal end of the balloon was not returned.The inner tubing is detached.The guidewire was returned with the catheter.The distal 80cm of the guidewire has been cut off.A section of the inner tubing is stretched over this section of the guidewire.A comparative catheter of the same size/catalog number was pulled and tested.The balloon was immersed in a body temperature bath and pressurized in 1 atm increments until the balloon failed.The rated burst pressure for this catalog number is 12 atm.This balloon did not burst until 20 atm, which is well above the labeled rated burst pressure.The balloon burst was caused by the over pressurization of the balloon.According to the report from the user facility, they took this balloon to 24 atm before it burst, which is double the labeled rated burst pressure for this catheter.They were also using this catheter off-label for stent restenosis.This catheter is indicated for angioplasty use only.There are warnings listed in the instructions for use about the over pressurization of the balloons - warnings: "caution: do not exceed the rbp.An inflation device with pressure gauge is recommended to monitor pressure.Pressure in excess of the rbp can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.".
 
Event or Problem Description
As reported by the distributor bis and the hospital: "the balloon burst.It is unknown whether an inflation device was used." updated 9/18/2018 - " we ballooned the innominate vein in stent restenosis with the mini ghost to 24 atms, which was used 1 time before to inflate the balloon.The sheath was a 6fr mullins sheath and it was not kinked.The patient had congenital anomalies.After using the mini ghost, we used 6fr and 7fr 75 mustangs.The balloon burst at 24atms.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI GHOST PTA CATHETER
Common Device Name
PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY CATHETER
Manufacturer (Section D)
NUMED, INC.
2880 main street
hopkinton NY 12965
MDR Report Key7893140
Report Number1318694-2018-00012
Device Sequence Number18874392
Product Code DQY
UDI-Device Identifier04046964341427
UDI-Public04046964341427
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K051343
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model Number507
Device Catalogue NumberMG150
Device Lot NumberMG-2509
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/31/2018
Initial Report FDA Received Date09/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-