Clinical evaluation performed on october 15, 2018 by medical affairs director total hip revision surgery occurred more than 1 year after primary implantation.According to the report, implants did not achieve sufficient fixation.The causes for such an adverse event may vary across a wide range: one of the possibilities is undersizing of the stem, often correlated with peculiar anatomical situations; another possibility is silent infection, but there is no evidence of this situation in this case.Failure of achieving satisfactory primary stability compromises the chances of a secure secondary fixation and results in component loosening.We have no reason to suspect a faulty device in this case.Preliminary investigation performed on october 15, 2018 by r&d product manager: update of preliminary investigation: the branch "provided" additional photos and confirmed that the scratches were done during the revision.The stem showed a coating almost totally absorbed, but in the proximal region it was still evident.A possible cause is an undersized "implanted" stem, but from the received images it was not possible to determine with certainty the root cause of the event.Visual inspection performed on (b)(6) 2018 by the r&d project manager: the stem lost its coating almost entirely, leaving only the more proximal part with some residual of coatings.Moreover, the distal tip appeared scratched and the stem showed scratches and holes caused by the revision: the two facts can be related, probably caused by a distal fixation and a difficult extraction during the revision surgery, but from the received part it is not possible to determine the root cause with certainty.
|