AVANOS MEDICAL INC. MIC-KEY* GASTRIC-JEJUNAL FEEDING TUBE KIT - 16 FR, 2.0 CM; DH EF BALLOON TUBES PRODUCTS
|
Back to Search Results |
|
| Model Number |
0270-16-2.0-30 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history record for the reported lot number, aa7205n18, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 18-sep-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported a cross-lumen breach occurred on the jejunal feeding tube.The internal septum was compromised, allowing jejunal feeding to exit the gastric skives.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: one sample device was returned.No product packaging was received.The tubing exhibited a very dark discoloration on the interior.The sample emitted a strong odor.The sample was examined under ambient light conditions.Slight tearing was observed at the second gastric skive.Feeding was simulated, using water, through the jejunal feed port.The water infused into the jejunal port exited the gastric skives.The area was examined under magnification.The internal septum which separates the two feed lumens exhibited a breach.The septum appeared somewhat lumpy.The complaint was confirmed.No root cause was identified.The cause for this incident could not be conclusively determined.All information reasonably known as of 16-nov-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|