BECTON DICKINSON MEDICAL SYSTEMS BD LUER-LOK¿ SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE; SYRINGE WITH NEEDLE Back to Search Results
Catalog Number 309644
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that a bd luer-lok¿ syringe with bd precisionglide¿ needle had white residue on it.There was no report of exposure, serious injury, or medical intervention.
 
Event or Problem Description
It was reported that a bd luer-lok¿ syringe with bd precisionglide¿ needle had white residue on it.There was no report of exposure, serious injury, or medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: three photos and one loose 10ml assembled syringe with needle was received and reported to be from batch #8131751 (p/n 309644).The sample was visually evaluated.The needle cannula was found to have a piece of white foreign matter on exterior.The foreign matter appears to be the same epoxy used in the needle assembly process.The foreign matter is level three in size, which is rejectable per product specification.A device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Potential root cause for the foreign matter defect is associated with the needle assembly process.No corrective actions are necessary based on the defective rate identified.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK¿ SYRINGE WITH BD PRECISIONGLIDE¿ NEEDLE
Common Device Name
SYRINGE WITH NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key7894456
Report Number1213809-2018-00618
Device Sequence Number11121221
Product Code FMF
UDI-Device Identifier30382903096443
UDI-Public30382903096443
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K980987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2023
Device Catalogue Number309644
Device Lot Number8131751
Initial Date Received by Manufacturer 08/31/2018
Supplement Date Received by Manufacturer08/31/2018
Initial Report FDA Received Date09/20/2018
Supplement Report FDA Received Date10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-