| Model Number |
189602000 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Spinal Column Injury (2081)
|
| Date of Event |
08/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device was not returned for evaluation.A review of the device history record could not be performed as the lot number is unknown.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that dr.(b)(6) was doing a revision on the patient on monday (b)(6) 2018, he implanted another rod with open 6.35 connectors.This implantation procedure was unilateral.Upon exploration, he saw the set screw was extremely loose.Once he gained better exposure, he saw the threaded area of the lateral connector splayed.After removing the lateral connector, he saw the base cracked.Dr.(b)(6) replaced the connector with a new one.(1896-92-050).
|
| |
|