• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CLAMP, SPINOUS PROCESS TALL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CLAMP, SPINOUS PROCESS TALL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734715
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.The tall spinous process clamp was returned to the manufacturer for analysis.Analysis found that the returned clamp was in very good condition with no apparent physical damage.The clamp was of the design that does not allow the attachment screw to be backed out completely and threads back into the base in order to close the clamp.Clamp functioned as designed.No failure found.Device manufacture date unavailable.
 
Event or Problem Description
Medtronic received information regarding a navigation system event having occurred outside of procedure.It was reported that the site had a broken tall, ¿new style¿ spine clamp.It was stated it had a stripped screw.No patient present.
 
Additional Manufacturer Narrative
Correction: manufacture date provided.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAMP, SPINOUS PROCESS TALL
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key7894734
Report Number1723170-2018-04881
Device Sequence Number19194915
Product Code OLO
UDI-Device Identifier00643169936300
UDI-Public00643169936300
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K131425
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9734715
Device Lot Number170425
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2017
Supplement Date Received by Manufacturer11/01/2017
Initial Report FDA Received Date09/20/2018
Supplement Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-