• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIATRS45AXT
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic mini gastric bypass, on stomach transection, the handle was able to fire a load initially without incident.While trying to perform the second firing, the knife blade and staples did not advance.The green button was activated, the ring handle kicked out, but then upon pulling the ring handle it would squeeze as per usual, but there was no activity within the staple load.The teeth did not engage the pushing mechanism.This was evident again upon testing post-operatively.No malfunction was apparent during the first successful firing of the stapler.A new product was opened and used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.Visual and functional evaluation of the reload found no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7894748
Report Number1219930-2018-05144
Device Sequence Number16629274
Product Code GDW
UDI-Device Identifier10884521190993
UDI-Public10884521190993
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K133938
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Model NumberEGIATRS45AXT
Device Catalogue NumberEGIATRS45AXT
Device Lot NumberN8E1080KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Initial Date Received by Manufacturer 08/30/2018
Supplement Date Received by Manufacturer10/20/2018
Initial Report FDA Received Date09/20/2018
Supplement Report FDA Received Date11/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-