| Catalog Number |
10494134 |
| Medical Device Problem Code |
Accessory Incompatible (1004)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint was raised to manufacturing who investigated the issue.Investigation concluded that this was an isolated incident caused by manual stacking of the cards during manufacturing.Due to this finding, manufacturing is ensuring that all cassettes that have not been automatically processed by the line will be checked by the visions system to ensure they are packaged correctly and an alert has been raised to ensure future cassettes are automatically processed.There have been no other complaints for this issue.
|
| |
|
Event or Problem Description
|
|
The customer reported that in the box of the reagent cards (pro12) used on the novus, some were assembled in the opposite direction so that the test pads were read by the device in the incorrect order.The customer stated that they received incorrect protein results that were reported to the physician.The lab technician re-ran the samples and got correct results and a corrected report was issued.Siemens requested the results but they have not been provided.There was no report of injury due to this event.
|
| |
|