Model Number FUSION COMPACT
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/01/2017
Event Type  malfunction  
Manufacturer Narrative
Axiem system was returned to the manufacturer for analysis.Analysis found that when the axiem was connected to a test system, it was functioning correctly.The hardware investigation found that the reported event was related to a hardware issue.Manufacture date was unavailable.
 
Event Description
Medtronic received information regarding a navigation system during a functional endoscopic sinus surgery (fess).It was reported that the system was not communicating with the axiem box.The system was displaying the red 8 on the side.Both systems were rebooted and reseated to each other with no resolution the system still displaying the red 8.Navigation and medtronic imaging was aborted but replacing the axiem box resolved the issue.The surgery was completed with less than an hour delay and no patient impact.
 
Manufacturer Narrative
Additional information: manufacture date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
jonathan wong
navigation customer quality
826 coal creek circle
louisville, CO 80027
7635267745
MDR Report Key7895847
MDR Text Key120968731
Report Number1723170-2018-04825
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age51 YR
Patient Weight52
-
-