DEPUY ORTHOPAEDICS, INC. 1818910  LCS COMPLETE M/B PAT POR STD; LCS COMPLETE : KNEE PATELLA Back to Search Results
Catalog Number 129409640
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Migration (4003)
Health Effect - Clinical Codes Pain (1994); Tissue Damage (2104); No Code Available (3191)
Date of Event 08/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Representing surgical intervention.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient reported a painful knee post patella resurfacing surgery.X-rays showed a tilted patella.Upon opening the knee metallosis was evident and the patella poly had popped out of the correct position and the pin had locked into a new position in the poly.This locked the patella poly into a new position where it was not free to move.The metal back was also damaged from being exposed to the femoral component and rubbing on it, causing the metallosis.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot :8608187.If information is obtained that was not available for the initial medwatch , a follow-up medwatch will be filed as appropriate. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS COMPLETE M/B PAT POR STD
Common Device Name
LCS COMPLETE : KNEE PATELLA
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7896359
Report Number1818910-2018-69872
Device Sequence Number19310156
Product Code NJL
UDI-Device Identifier10603295022640
UDI-Public10603295022640
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number129409640
Device Lot Number8608187
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer11/13/2018
Initial Report FDA Received Date09/20/2018
Supplement Report FDA Received Date11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-