| Model Number |
MMT-723RNAS |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 135 mg/dl.The customer stated that the drive support cap was protruded.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device alarmed motor error during the basic occlusion test due to loose drive support disk.Unable to perform basic occlusion test, occlusion test,compromised forced sensor system test, excessive no delivery test, displacement test or verify compromised forced sensor system alarm due to motor error alarm.However, device passed rewind test and displacement test.
|
| |
|
Search Alerts/Recalls
|