| Model Number |
LX AQUA CALIBRATOR LEVEL 1 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A beckman coulter applications specialist was sent to the customer site.A new lot of aquacal level 1 was used to resolve the issue.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported erroneously low sodium (na) results on their unicel dxc 600 pro synchron system.The results were reported outside of the laboratory.The customer stated that quality control (qc) recovered within specifications prior to the event.There was no change to patient treatment in association with this event.
|
| |
|