MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FUSE; ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FUSE; ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT Back to Search Results
Model Number FSC-3300-ST
Medical Device Problem Codes Image Display Error/Artifact (1304); Poor Quality Image (1408)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the center screen had horizontal lines.The anatomy was not visible.A second fuse 1c colonoscope was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Device code 1304 captures the reportable issue of center image lost.A fuse 1c colonoscope was received for analysis.A functional evaluation was performed and the reported issue of "line across middle screen" could not be replicated.The repair history was reviewed and nothing was found to indicate a possible service-related cause for the complaint.Since the reported issue could not be replicated, the assigned investigation conclusion code for this complaint is designated as no problem detected.This supplemental correction report is being filed to correct the report number in a previously submitted supplemental report (follow-up number 001) which was accepted by fda on 12/27/2018 01:30 pm ct.The report number is being corrected from: 3005099803-2018-62283 to: 3005099803-2018-60489.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the center screen had horizontal lines.The anatomy was not visible.A second fuse 1c colonoscope was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSE
Common Device Name
ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key7897335
Report Number3005099803-2018-60489
Device Sequence Number18874395
Product Code PEA
UDI-Device Identifier00816849019666
UDI-Public00816849019666
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K160275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFSC-3300-ST
Device Catalogue Number60093
Device Lot NumberF1C0924E1C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2018
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer12/03/2018
Initial Report FDA Received Date09/21/2018
Supplement Report FDA Received Date01/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-