MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 20188
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event was approximated to (b)(6) 2018 as no event date was reported.
 
Event or Problem Description
It was reported that stent dislodgment occurred.During preparation of a 7.0mmx19mmx90cm express sd renal/biliary stent, upon insertion into the sheath, the stent came off the stent delivery balloon.The device never entered the patient's body.No patient complications were reported.
 
Event or Problem Description
It was reported that stent dislodgment occurred.During preparation of a 7.0mmx19mmx90cm express sd renal/biliary stent, upon insertion into the sheath, the stent came off the stent delivery balloon.The device never entered the patient's body.No patient complications were reported.The returned product consisted of the complaint device, mandrel and balloon protector.Visual and microscopic analysis on the outer shaft, inner shaft, balloon and tip noted that stent is flattened and damaged.The balloon was still tightly folded and the stent impression marks when it was cramped on the balloon is still visible.There is blood present in the guidewire lumen, on the stent and on the balloon folds.Inspection of the remainder of the device presented no other damage or irregularities.
 
Additional Manufacturer Narrative
Date of event was approximated to (b)(6) 2018 as no event date was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key7897559
Report Number2134265-2018-61180
Device Sequence Number11111529
Product Code NIN
UDI-Device Identifier08714729390527
UDI-Public08714729390527
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K152607
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2018
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0018019499
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Initial Date Received by Manufacturer 08/29/2018
Supplement Date Received by Manufacturer09/28/2018
Initial Report FDA Received Date09/21/2018
Supplement Report FDA Received Date10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-