Model Number G28781
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/24/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This mdr is being filed after the associated complaint was reviewed under remediation protocol cap (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
The complainant reported that the procedure was for iliac stenting from a contralateral puncture site, but the end of the amplatz stiff support wire guide was ¿very floppy.¿ it was further noted that the anchor solder was separated from the inner mandrel wire which resulted in coil elongation.The product problem did not result in any reported serious injuries to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZ STIFF SUPPORT WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7897948
Report Number1820334-2018-02894
Device Sequence Number1807104
Product Code DQX
UDI-Device Identifier00827002287815
UDI-Public(01)00827002287815(17)200513(10)5803539
Combination Product (Y/N)N
Initial Reporter CountryGB
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/13/2020
Device Model NumberG28781
Device Catalogue NumberTHSF-35-180-ASG
Device Lot Number5803539
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2018
Initial Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-