Catalog Number 0066214
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Immucor technical support used a remote electronic connection method on 01sep2018 to assess the instrument test well images in question, and determined that they appeared visually negative.No errors were noted in the instrument event log.The immucor laboratory tested retention product on 14sep2018, which performed as expected.The immucor laboratory also received a blood sample from the customer site ((b)(4)) and tested it on 14sep2018 against retention product of id363, which yielded the same unexpectedly negative outcome as the customer had obtained.The immucor laboratory also tested that same returned blood sample with antibody screen lot number r988, which also yielded an unexpectedly negative outcome.An immucor field service engineer (fse) visited the customer site on 04sep2018 to assess the echo lumena testing instrument and determined that the instrument was operating as expected.(b)(4).
 
Event or Problem Description
On (b)(6) 2018, a customer site reported an unexpectedly negative antibody identification when using capture-r ready-id, when used on an echo lumena instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-ID
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key7898032
Report Number1034569-2018-00211
Device Sequence Number18903319
Product Code KSZ
UDI-Device Identifier10888234001188
UDI-Public10888234001188
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
102707/0.0
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date09/18/2018
Device Catalogue Number0066214
Device Lot NumberID363
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2018
Initial Report FDA Received Date09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
Patient Weight82
-
-