• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VASCUTRAK PTA DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VASCUTRAK PTA DILATATION CATHETER Back to Search Results
Catalog Number V1850150
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Needle Stick/Puncture (2462)
Date of Event 06/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the femoral artery, the pta balloon allegedly would not deflate.A needle was inserted through the skin to deflate the balloon.Another device was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the femoral artery, the pta balloon allegedly would not deflate.A needle was inserted through the skin to deflate the balloon.Another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The device and three electronic radiographic images were returned for evaluation.No visual defects were noted to the catheter shaft or inflation luer.The balloon was inflated with an in-house inflation device and water was seen exiting the balloon from the user-induced puncture site in the balloon.Therefore, the investigation is inconclusive for the reported deflation issue, as the balloon was returned punctured and could not be fully functionally tested.The review of the images did not confirmed the reported deflation issue.Based on the evaluation results, the device was unable to be fully functionally tested due to the puncture in the balloon.Therefore, the definitive root cause for the reported deflation issues could not be determined.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASCUTRAK PTA DILATATION CATHETER
Common Device Name
PTA DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7898606
Report Number2020394-2018-01759
Device Sequence Number10805129
Product Code PNO
UDI-Device Identifier00801741061691
UDI-Public(01)00801741061691
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K103459
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue NumberV1850150
Device Lot NumberGFCP3856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2018
Initial Date Received by Manufacturer 08/27/2018
Supplement Date Received by Manufacturer10/19/2018
Initial Report FDA Received Date09/21/2018
Supplement Report FDA Received Date11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-