| Model Number |
ZCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
|
| Date of Event |
08/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, the lens was removed/replaced within the initial procedure.If explanted, give date: not applicable, the lens was removed/replaced within the initial procedure.(b)(4).Device evaluation: the product was not returned to the manufacturing site; therefore, product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: during the manufacturing record review found no product deficiency identified.The lens was manufactured according to specifications.A review of complaints revealed that no similar complaints were received for this production order number.Labeling review: the labeling review was completed and revealed that the directions for use (dfu) provide the customer with proper usage instructions and guidelines.As a result of the investigation, there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient¿s capsular bag broke when inserting the zcb00 25.5 diopter intraocular lens (iol) into the left eye (os).Per surgeon¿s operative report, the patient's cataract was noted to have very sharp edges that were adherent to the capsule itself.The one piece lens was placed and was noted to be very mobile and displaced posteriorly.The replacement lens was a non j&j intraocular lens.An incision enlargement, vitrectomy, and sutures were required due to the complications of the capsule tearing.Patient was doing fine post op.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint device was not received at the time the initial report was filed on 09/21/2018.The device was received at the manufacturing site on 10/1/2018 where product testing has been completed.Device available for evaluation? yes.Returned to manufacturer on: 10/1/2018.Device returned to manufacturer? yes.Device evaluation: visual inspection of the returned lens showed a scratch at the optic area and one of the haptics was broken; most probably to make removal possible.Based on the visual evaluation of the returned unit, it could not be determined that the cause of the issue reported is related to the manufacturing process.A product quality deficiency could not be confirmed.The customer's reported event could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|